      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64127",
      "recalling_firm": "Philips Healthcare Inc.",
      "address_1": "3000 Minuteman Road",
      "address_2": "N/A",
      "postal_code": "01810",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution including the states of FL, CA, WA, AZ, NC, KY, DE, AZ and MO.",
      "recall_number": "Z-0767-2013",
      "product_description": "Phillips Easy Diagnost Eleva DRF. Model Number: 706034    Product Usage:  Universal Fluoroscopic Applications",
      "product_quantity": "14 units",
      "reason_for_recall": "The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system.",
      "recall_initiation_date": "20130116",
      "center_classification_date": "20130201",
      "termination_date": "20170123",
      "report_date": "20130213",
      "code_info": "Serial Numbers 362600, 362601, 364209, 364210, 375245, 375850, 376439, 384396, 387492, 398185, 399440, 399621, 399623, 402409",
      "more_code_info": ""
    }
  ]
}