{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Austin",
      "address_1": "9800 Metric Blvd",
      "reason_for_recall": "It has been discovered that during Empowr PS Total Knee replacement, while chiseling for bone removal in the box area, femoral chipping has occurred in surgery.",
      "address_2": "",
      "product_quantity": "162 units",
      "code_info": "202351L01, 202351L02, 202351L03, 202351L04, 202351L05, 202351L06, 202351L07, 202351L08, 202351L09, 203779L09, 202351L10, 203779L10, 203203L01",
      "center_classification_date": "20161214",
      "distribution_pattern": "Nationwide Distribution",
      "state": "TX",
      "product_description": "Empowr PS Knee System Box Cut Guide  The EMPOWER PS Box Cut Guide is used to make the housing resection for the post of the posterior stabilized implant",
      "report_date": "20161221",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Encore Medical, Lp",
      "recall_number": "Z-0766-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75645",
      "termination_date": "20171020",
      "more_code_info": "",
      "recall_initiation_date": "20160919",
      "postal_code": "78758-5445",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}