      "status": "Terminated",
      "city": "Glendora",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64163",
      "recalling_firm": "Ormco/Sybronendo",
      "address_1": "1332 S Lone Hill Ave",
      "address_2": "N/A",
      "postal_code": "91740-5339",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA and the country of Canada.",
      "recall_number": "Z-0765-2013",
      "product_description": "Sybron Endo Lightspeed: LSX Files, 25MM, Part Numbers: 835-2525; LSX Files, 25MM Assorted Kit, Part Numbers: 835-2455; LSX Files, 30MM, Part Numbers: 835-2031.     An engine-driven rotary endodontic instrument used during endodontic procedures for cleaning and shaping the root canal.",
      "product_quantity": "740 units",
      "reason_for_recall": "The length markings on the Lightspeed LSX files were wrong.",
      "recall_initiation_date": "20121212",
      "center_classification_date": "20130201",
      "termination_date": "20131107",
      "report_date": "20130213",
      "code_info": "Part # 835-2525, Lot # 091274189.  Part # 835-2455, Lot # 091272054, 091279295, 101200609, 101200610, 101200612.  Part # 835-2031, Lot # 091268204.",
      "more_code_info": ""
    }
  ]
}