  "results": [
    {
      "status": "Ongoing",
      "city": "Buccinasco",
      "state": "N/A",
      "country": "Italy",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84309",
      "recalling_firm": "Villa Sistemi Medicali S.P.A.",
      "address_1": "Via delle Azalee 3",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, OH, OK, PA, TN, TX, UT, and Puerto Rico.",
      "recall_number": "Z-0764-2020",
      "product_description": "Apollo, Model Numbers 9784100231, 9784120131, 9784130131 - Product Usage: Apollo table is a remote controlled X-ray system suitable for radiography and fluoroscopy examinations. The system is stationary mounted. The movements of the table and the tube stand are motorized and remotely controlled. Medical purpose is the diagnosis, visualization and documentation of various human diseases or injuries.",
      "product_quantity": "3",
      "reason_for_recall": "An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit.  After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.",
      "recall_initiation_date": "20191104",
      "center_classification_date": "20191231",
      "report_date": "20200108",
      "code_info": "Serial Numbers: 08020473  11100873  15041446"
    }
  ]
}