  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72571",
      "recalling_firm": "Philips Electronics North  America Corporation",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution.  US states of TN and IN and the countries of Australia, Belgium, Egypt, France, Germany, Japan, Republic of Korea, Netherlands, and Saudi Arabia.",
      "recall_number": "Z-0764-2016",
      "product_description": "Philips Allura Xper FD20 Ceiling version only; 722028; imaging applications.",
      "product_quantity": "15 subject to correction and removal",
      "reason_for_recall": "Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is initially stopped in an unbalanced position, the air gap of the brake can become too large, resulting in continued C-arc movement (C-arc is looking for balanced position).",
      "recall_initiation_date": "20151103",
      "center_classification_date": "20160205",
      "termination_date": "20160819",
      "report_date": "20160217",
      "code_info": "Ceiling version only.",
      "more_code_info": ""
    }
  ]
}