  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Saint Paul",
      "address_1": "177 E County Road B",
      "reason_for_recall": "The distal end of the core wire of the TorqVue FX Delivery System could potentially fracture when exposed to a combination of certain cardiac anatomies and usage conditions.",
      "address_2": "",
      "product_quantity": "635",
      "code_info": "all batches",
      "center_classification_date": "20130211",
      "distribution_pattern": "Nationwide Distribution-including DC and the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TX, VA, WA, and WI.",
      "state": "MN",
      "product_description": "St. Jude Medical, AGA Medical corporation, AMPLATZER¿ TorqVue¿ FX Delivery System. Model #'s: 9-ITVFX06F45/60, 9-ITVFX07F45/60, 9-ITVFX007F45/80, 9-ITVFX08F45/60, 9-ITVFX08F45/80, 9-ITVFX09F45/80, 9-ITVFX10F45/80 9-ITVFX12F45/80, 9-ITVFX13F45/80. Cardiac surgical instrument, used to close atrial septal defects between the right and left atriums.",
      "report_date": "20130220",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "St Jude Medical Inc",
      "recall_number": "Z-0763-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64155",
      "termination_date": "20130521",
      "more_code_info": "",
      "recall_initiation_date": "20130118",
      "postal_code": "55117-1951",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}