  "results": [
    {
      "status": "Terminated",
      "city": "Covington",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72756",
      "recalling_firm": "C.R. Bard, Inc.",
      "address_1": "8195 Industrial Blvd NE",
      "address_2": "N/A",
      "postal_code": "30014-1497",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution-DC and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI. IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.",
      "recall_number": "Z-0760-2016",
      "product_description": "Arctic Sun 5000 and Arctic Sun 2000;  Intended for monitoring and controlling patient temperatures.",
      "product_quantity": "1,539",
      "reason_for_recall": "Improper targeted temperature therapy was delivered to patients because the Temperature Simulator Key was connected to the patient temperature cable instead of the patient temperature probe.",
      "recall_initiation_date": "20151110",
      "center_classification_date": "20160205",
      "termination_date": "20170421",
      "report_date": "20160217",
      "code_info": "All lots/serial numbers sold prior to August 2014.",
      "more_code_info": ""
    }
  ]
}