  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Charleston",
      "state": "SC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73103",
      "recalling_firm": "Trumpf Medical Systems, Inc.",
      "address_1": "1046 Legrand Blvd",
      "address_2": "N/A",
      "postal_code": "29492-7672",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US including MA and TX; Foreign distribution to Austria, Azerbaijan, Belgium, British Virgin Islands, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Luxemburg, Morocco, Netherlands, New Caledonia, Norway, Poland, Portugal, Qatar, Romania, San Marino, Spain, Sweden and Switzerland.",
      "recall_number": "Z-0759-2016",
      "product_description": "Patient transport Shuttle with Trendelenburg quick adjustment function (Jupiter, Mars, Titan, TruSystem and Saturn operating table systems).  The shuttle is intended for the following applications: 1. Transfer, transport and storage of exchangeable table tops in the Trumpf Medical TruSystem 7500, JUPITER, and SATURN SM operating table systems 2. Patient transport to an operating table top (without a column) from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system 3. Transporting the TruSystem 7500 SM, JUPITER SM or SATURN SM operating table column with table top (without a patient) within the operating theatre.",
      "product_quantity": "1284 affected units",
      "reason_for_recall": "Complaints were received of  the table top on the transporter (shuttle) tilting down quickly in a maximum Trendelenburg position.",
      "recall_initiation_date": "20160119",
      "center_classification_date": "20160205",
      "termination_date": "20160524",
      "report_date": "20160217",
      "code_info": "Product: Shuttle 2.7, Material Number: 1254362, Serial Number Range: 100801175 \u0013 101349420; Product: Shuttle 2.7 360, Material Number: 1459862, Serial Number Range: 100843669-101212847; Product: Shuttle 3.7, Material Number: 1254363, Serial Number Range: 100180383-101290987 and Product: Shuttle 3.7 360, Material Number: 1459863, Serial Number Range: 100787418-101334652.",
      "more_code_info": ""
    }
  ]
}