  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84034",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US  OUS: Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Korea (Republic of), Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Peru, Poland, Republic of Serbia, Romania, Russia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uganda, and Unites Kingdom",
      "recall_number": "Z-0757-2020",
      "product_description": "CARESCAPE Respiratory Module E-sCAiOV",
      "product_quantity": "37 units",
      "reason_for_recall": "Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.",
      "recall_initiation_date": "20190927",
      "center_classification_date": "20200110",
      "termination_date": "20230926",
      "report_date": "20200122",
      "code_info": "Mfg. Lot or Serial # UDI #  SGV19283064HA 010084068210414221SGV19283064HA11190710  SGV19283042HA 010084068210414221SGV19283042HA11190710  SGV19243024HA 010084068210414221SGV19243024HA11190610  SGV19283053HA 010084068210414221SGV19283053HA11190710  SGV19283001HA 010084068210414221SGV19283001HA11190710  SGV19283002HA 010084068210414221SGV19283002HA11190710  SGV19283004HA 010084068210414221SGV19283004HA11190710  SGV19283006HA 010084068210414221SGV19283006HA11190710  SGV19283007HA 010084068210414221SGV19283007HA11190710  SGV19283008HA 010084068210414221SGV19283008HA11190710  SGV19283009HA 010084068210414221SGV19283009HA11190710  SGV19283011HA 010084068210414221SGV19283011HA11190710  SGV19283012HA 010084068210414221SGV19283012HA11190710  SGV19283013HA 010084068210414221SGV19283013HA11190710  SGV19283015HA 010084068210414221SGV19283015HA11190710  SGV19283016HA 010084068210414221SGV19283016HA11190710  SGV19283017HA 010084068210414221SGV19283017HA11190710  SGV19283019HA 010084068210414221SGV19283019HA11190710  SGV19283024HA 010084068210414221SGV19283024HA11190710  SGV19283025HA 010084068210414221SGV19283025HA11190710  SGV19283032HA 010084068210414221SGV19283032HA11190710  SGV19283033HA 010084068210414221SGV19283033HA11190710  SGV19283034HA 010084068210414221SGV19283034HA11190710  SGV19283035HA 010084068210414221SGV19283035HA11190710  SGV19283037HA 010084068210414221SGV19283037HA11190710  SGV19283038HA 010084068210414221SGV19283038HA11190710  SGV19283044HA 010084068210414221SGV19283044HA11190710  SGV19283046HA 010084068210414221SGV19283046HA11190710  SGV19283047HA 010084068210414221SGV19283047HA11190710  SGV19283050HA 010084068210414221SGV19283050HA11190710  SGV19283058HA 010084068210414221SGV19283058HA11190710  SGV19283059HA 010084068210414221SGV19283059HA11190710  SGV19283060HA 010084068210414221SGV19283060HA11190710  SGV19283061HA 010084068210414221SGV19283061HA11190710  SGV19283065HA 010084068210414221SGV19283065HA11190710  SGV19283070HA 010084068210414221SGV19283070HA11190710  SGV17333012HA Not Applicable",
      "more_code_info": ""
    }
  ]
}