  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69838",
      "recalling_firm": "Stryker Endoscopy",
      "address_1": "5900 Optical Ct",
      "address_2": "N/A",
      "postal_code": "95138-1400",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.",
      "recall_number": "Z-0757-2015",
      "product_description": "PKG, 5MM PEEK HANDLE, 45CM, P/N 0250181164  Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.",
      "product_quantity": "180,573 units total",
      "reason_for_recall": "The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods",
      "recall_initiation_date": "20141117",
      "center_classification_date": "20141217",
      "termination_date": "20150717",
      "report_date": "20141224",
      "code_info": "01292013D  0247065C  0341845C  0342941C  0345124C  0345125C  4082014  0540391C  0540392CD  0542893D  0647218D  07122013E  0743480D  0744062D  0843047D  0845027D  0845859D  0846783D  0940144D  0940148D  0940821D  0941601D  0943110D  0943251D  0943541D  0944721D  0945231D  0945465D  1040630D  1040806D  1041063D  1041468D  1042102D  1043037D  1043039D  1044650D  1045037D  1045055D  1045227D  1045481D  1045638D  1045841D  1046702D  11082010A  1141562  1141562D  1141969D  1141970D  1142168D  1143113D  1143150D  1144236D  1144341D  1144655D  1145341D  1145632D  1145809D  1145939  1145939D  1146669D  1146671D  12122013  1241797D  1241938D  1243268D  1243269D  1244956  1244956D  1245522  1245798  1245799  1245799H  1245800  1245977  1340938  1341257  1341872  1343540  1343540H  1343541  1343542  1344000  1344000H  1344251  1440346  1440346H  1440347  1440347H  1440792  1442304  1442799"
    }
  ]
}