  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waukegan",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93628",
      "recalling_firm": "Cardinal Health 200, LLC",
      "address_1": "3651 Birchwood Dr",
      "address_2": "N/A",
      "postal_code": "60085-8337",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US, Canada, Australia and New Zealand",
      "recall_number": "Z-0755-2024",
      "product_description": "Cardinal Health  MEDI-VAC CRD Suction Canister Liner, 1000 mL Suction Canister Liner Kit with 5 mm 1.8 m Tube, REF 65651-516",
      "product_quantity": "203244 units",
      "reason_for_recall": "The above-listed lots of Cardinal Health Medi-Vac CRD\" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.",
      "recall_initiation_date": "20231221",
      "center_classification_date": "20240118",
      "report_date": "20240124",
      "code_info": "50630140152633 (case), Lot Numbers:  J308-606, J308-624, J308-628 J308-642, J308-643, J309-608, J309-610, J309-622, J309-626, J309-633, J309-645, J310-602, J310-617, J310-645"
    }
  ]
}