{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Rochester",
      "address_1": "100 Indigo Creek Dr",
      "reason_for_recall": "Some VITROS¿ Chemistry Products Vapor Adsorption Cartridge CAT No. 6800100 cartons incorrectly contain a Particulate Cartridge Part No. J19612. In some instances, the Particulate Cartridge has been inadvertently installed on the VITROS¿ System instead of the intended VITROS¿ Vapor Adsorption Cartridge which may bias results.",
      "address_2": "",
      "product_quantity": "Domestic: 15,891; Foreign: 5237",
      "code_info": "CAT No. 680 0100; the cartridges do not have lot numbers.",
      "center_classification_date": "20130130",
      "distribution_pattern": "Worldwide Distribution, including Nationwide (USA) and the countries of AU, GM, FR, IT, SP, BR, BM, AR, CA, JA, MX, UK, CH, CL, CO, IN, EC, PE, PY, PA, PT, NZ, SG, VE and UY.",
      "state": "NY",
      "product_description": "Ortho Clinical Diagnostics VITROS¿ Chemistry Products Vapor Adsorption Cartridge  REF/CAT No. 680 0100.  VITROS¿ Vapor Adsorption Cartridges are used to remove vapor contaminants that may potentially interfere with Signal Reagent used with MicroWell Assays on the VITROS¿ 3600, VITROS¿ 5600, and VITROS¿ ECi/ECiQ Systems.",
      "report_date": "20130206",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "recall_number": "Z-0755-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "63934",
      "termination_date": "20180705",
      "more_code_info": "",
      "recall_initiation_date": "20121213",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}