  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "East Syracuse",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86800",
      "recalling_firm": "Intersurgical Inc",
      "address_1": "6757 Kinne St",
      "address_2": "N/A",
      "postal_code": "13057-1215",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "All of the devices were shipped to the same distributor (McKesson).",
      "recall_number": "Z-0753-2021",
      "product_description": "Smoothbore breathing system with exhalation port 22 mm Device REF #: B0062001 Custom Anasthesia Breathing Systems; Multipac Anasthesia Breathing Systems - Product Usage: intended to administer medical gases to a patient during anesthesia. It provides both an inhalation and exhalation route and may include a connector, adaptor, and   Y-piece.",
      "product_quantity": "200 devices",
      "reason_for_recall": "The device was assembled with the incorrect exhalation port which can cause gas leakage.",
      "recall_initiation_date": "20201111",
      "center_classification_date": "20201230",
      "termination_date": "20220809",
      "report_date": "20210106",
      "code_info": "Model Number: B0062001  Lot Numbers: 4200283 and 4200286  UDI Codes:  (01)00081647202239(10)4200286  (01)00081647202239(10)4200283"
    }
  ]
}