  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rillieux La Pape",
      "state": "N/A",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84273",
      "recalling_firm": "Medicrea International",
      "address_1": "5389 Route De Strasbourg",
      "address_2": "Vancia",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US: NY, CO, OH, AZ, FL, TX, PA  OUS: France, Belgium, South Africa, China, Spain, Portugal",
      "recall_number": "Z-0753-2020",
      "product_description": "Medicrea Pass LP, REF B02217545, Non Sterile, Material: Ti-6Al-4V ELI, VIS PEDICULAIRE POLYAXIALE 7.5MM x 45MM, POLYAXIAL PEDICLE SCREW 7.5MM x 45 MM, POLYAXIAL PEDIKELSCHRAUBE 7.5MM X 45MM, TORNILLO PEDICULAR POLIAXIAL 7.5MM x 45MM  VITE POLIASSIALE PEDUNCOLARE 7.5MM x 45MM, PARAFUSO PEDICULAR POLIAXIAL 7.5MM x 45MM, POLYAXIALE PEDICULAIRE SCHROEF 7.5MM x 45MM, POLYAXIALE PEDICULAIRE SCHROEF 7.5MM x 45MM, WIELOOSIOWA SRUBA TRANSPEDIKULARNA 7.5MM x 45MM, (01)03613720177906(10)19A0630(17)290103, RX only",
      "product_quantity": "210 Pedicle Screws",
      "reason_for_recall": "It has been determined that some Pedicle system screw  identified as 45mm long have an actual length of 50mm. Pedicle screws with a diameter of 7.5 mm or more are reserved for the sacrum area. As a result, a screw that is too long could come into contact with the nerves and cause postoperative pain.",
      "recall_initiation_date": "20190627",
      "center_classification_date": "20191230",
      "termination_date": "20201119",
      "report_date": "20200108",
      "code_info": "Lot #/Expiration Date:  19A0630/January 03, 2029  19B0329 /January 21th, 2029  19C0734 /March 03, 2029"
    }
  ]
}