      "status": "Terminated",
      "city": "Morrisville",
      "state": "NC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81909",
      "recalling_firm": "Teleflex Medical",
      "address_1": "3015 Carrington Mill Blvd",
      "address_2": "N/A",
      "postal_code": "27560-5437",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of:  Australia, Belgium, China, Czech Republic, France, Germany, Great Britain, Iran, Ireland, Italy, Japan, Lebanon, Netherlands, Qatar, Romania, Saudi Arabia, Singapore, Spain, Taiwan, Turkey, and the United Arab Emirates.",
      "recall_number": "Z-0753-2019",
      "product_description": "Percuvance(R) Percutaneous Shaft 29cm, PCVSH3    The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate Intended Use: soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-o-lok ligating clips to soft tissue during laparoscopic surgery.",
      "product_quantity": "1751 units",
      "reason_for_recall": "There is potential for a piece of the distal tip on the Percuvance shaft to become deformed and possibly break off during use. If a piece did break off it could compromise the functionality and security of the tool tip. Should this occur during use, surgical intervention may be required.",
      "recall_initiation_date": "20180103",
      "center_classification_date": "20190118",
      "termination_date": "20201030",
      "report_date": "20190130",
      "code_info": "Batch/Lot Numbers:  73A1700531, 73B1700180, 73F1700583, 73F1700584,   73F1700585, 73H1700728, 73J1600394, 73J1700072, 73J1700073, 73J1700074, 73J1700554"
    }
  ]
}