  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Parsippany",
      "address_1": "300 Interpace Pkwy Ste 410A",
      "reason_for_recall": "Lag screw measurement is short (approximately 5 mm) when using current screw measuring device (p/n 101-00010) per the IO FIX Surgical Technique Rev C. which can lead to possible inappropriate implant size selection (e.g. shorter screw size selected).",
      "address_2": "",
      "product_quantity": "93 units",
      "code_info": "Catalog Number 101-00010  Lot number    AP3V12  AP4E01  AP4N04  AP5E25  AP6A44  AP6E42  AP6L14  AP6M17  AP6T08",
      "center_classification_date": "20130130",
      "distribution_pattern": "Worldwide Distribution -- Nationwide Distribution including the states of AR, AZ, CA, FL, GA, IL, IN, MD, MT, NC, NE, NH, NJ, NY, OH, PA, TN, TX, WI and WA.",
      "state": "NJ",
      "product_description": "Extremity Medical, IO FiX System Accessory.  Catalog Number 101-00010.      This screw measuring device is an accessory to the IO FiX Screw and Washer System that is used to measure the length of the screw by placing the depth gauge over the guidewire and down to the bone. Orthopedic device.",
      "report_date": "20130206",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Extremity Medical LLC",
      "recall_number": "Z-0752-2013",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "63805",
      "termination_date": "20130613",
      "more_code_info": "",
      "recall_initiation_date": "20120322",
      "postal_code": "07054-1148",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}