{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Waukesha",
      "address_1": "3000 N Grandview Blvd",
      "reason_for_recall": "Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.",
      "address_2": "",
      "product_quantity": "276 units total, 109 of those nationwide.",
      "code_info": "Models M1139417, M1223628, M1192226-072951, M1192226-084200. M1192226-095181, M1192226-104599",
      "center_classification_date": "20130208",
      "distribution_pattern": "Worldwide Distribution - Worldwide Distribution - USA including AL, AZ, CA, CO, FL,GA, IA, ID, IL, KY, MA,  MD, MI, MN, MO, MS, MT, NC, ND ,NE, NM, NV, NY, OH, OR, PA ,TX,  VT, WA, WI and WV; and the countries of Australia, Belgium, Canada, China, France, New Zealand, Poland, Saudi Arabia, Switzerland and United Kingdom.",
      "state": "WI",
      "product_description": "GE Healthcare - Giraffe and Panda T-Piece Resuscitation System    The T-piece Resuscitation System [ventilator, emergency powered resuscitator]/Panda Bag and Mask Resuscitation System provides the basic equipment required for pulmonary resuscitation of infants.",
      "report_date": "20130220",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "GE Healthcare, LLC",
      "recall_number": "Z-0748-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "63761",
      "termination_date": "20170118",
      "more_code_info": "",
      "recall_initiation_date": "20121015",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}