  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Hopkinton",
      "address_1": "68 Elm St",
      "reason_for_recall": "Filter Tips molded incorrectly and may not seal to the Varispan tip adaptor properly causing inaccurate aspiration and dispensing",
      "address_2": "",
      "product_quantity": "595 trays (571,200 tips)",
      "code_info": "Lot Numbers:15470394, 15471775, 1601067, 16071147",
      "center_classification_date": "20161212",
      "distribution_pattern": "Worldwide Distribution - US including CA, MA, NY, IA, OR , PA, WA and Internationally  to Belgium, Canada, Finland, France, Germany, Japan, Netherlands, Singapore, and UK",
      "state": "MA",
      "product_description": "Perkin Elmer  RoboRack 25ul, Clear, Non-Conductive, Filter Tips, Pre-sterilized   Item Number: 6000689.  Supplied for use with JANUS Automated Workstations:  Product Number: AJS4001 AJM4001 AJI4001, AJL4001, AJS8001 , AJM8001, AJI8001, AJL8001, AJM4G01 , AJI4G01 , AJL4G01, AJM8G01, AJI8G01, AJL8G01, AJM4M01, AJI4M01 , AJL4M01, AJM8M01, AJI8M01, AJL8M01, AGS4NGS  The JANUS Automated Workstation is an automated, programmable liquid handling instrument intended for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation work flows. The NGS Express is an automated, programmable liquid handling instrument suitable for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation workflows",
      "report_date": "20161221",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "PerkinElmer, Inc",
      "recall_number": "Z-0747-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75508",
      "termination_date": "20180731",
      "more_code_info": "",
      "recall_initiation_date": "20160624",
      "postal_code": "01748-1602",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}