  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 Gbc Dr Ms 514",
      "reason_for_recall": "Siemens Healthcare Diagnostics has received complaints of IRON calibrations outside of acceptable calibration slope (m) values (0.97 - 1.03) when using IRON calibrator lots 2DN060 and 2FN063.  Lots 2GN068 and 2JN074 may also be impacted but no complaints have been received.",
      "address_2": "PO BOX 6101",
      "product_quantity": "5999",
      "code_info": "Siemens material number 10445010 -- and lot numbers 2DN060, exp 2013-10-01; 2FN063, exp 2013-12-01; 2GN068, exp 2014-01-01; and 2JN073, exp 2014-03-01.",
      "center_classification_date": "20130128",
      "distribution_pattern": "Nationwide Distribution.",
      "state": "DE",
      "product_description": "Dimension(R) IRON Calibrator (DC85), Siemens material number 10445010.    The Dimension(R) IRON Calibrator is an in vitro diagnostic product intended to be used to calibrate the Iron method for the Dimension(R) clinical chemistry system.",
      "report_date": "20130206",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-0745-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64081",
      "termination_date": "20140108",
      "more_code_info": "",
      "recall_initiation_date": "20121220",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}