  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cambridge",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89602",
      "recalling_firm": "Cytocell Ltd.",
      "address_1": "418 Cambridge Science Park, Milton Road",
      "address_2": "Oxford Gene Technology",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the states of Hawaii, Missouri, Pennsylvania, Tennessee.",
      "recall_number": "Z-0741-2022",
      "product_description": "Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS  Catalog Number: USA-LPH024",
      "product_quantity": "19 kits",
      "reason_for_recall": "individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result",
      "recall_initiation_date": "20220118",
      "center_classification_date": "20220309",
      "termination_date": "20241108",
      "report_date": "20220316",
      "code_info": "Lot Number: 075711 Exp. Date: 2023-06-22  UDI: (01)05055844901551(17)230622(10)075 711"
    }
  ]
}