{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Irvine",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73033",
      "recalling_firm": "Horiba Instruments Inc",
      "address_1": "9755 Research Dr",
      "address_2": "N/A",
      "postal_code": "92618-4626",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN    No foreign distribution.",
      "recall_number": "Z-0740-2016",
      "product_description": "ABX PENTRA Glucose PAP CP ref. A11A01668  reagent is intended for the quantitative in vitro diagnostic determination of glucose in human serum, plasma and urine using glucose oxidase method by colorimetry.",
      "product_quantity": "216",
      "reason_for_recall": "N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.",
      "recall_initiation_date": "20160111",
      "center_classification_date": "20160202",
      "termination_date": "20160607",
      "report_date": "20160210",
      "code_info": "All Lots",
      "more_code_info": ""
    }
  ]
}