  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Parsippany",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84402",
      "recalling_firm": "Diagnostica Stago, Inc.",
      "address_1": "5 Century Dr",
      "address_2": "",
      "postal_code": "07054-4607",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: MA ME VT CT NJ NY PA DC MD VA NC SC GA FL AL TN MS KY OH IN MI IA WI MN SD ND MT IL MO KS LA AR OK TX CO AZ NV CA HI OR WA AK",
      "recall_number": "Z-0738-2020",
      "product_description": "STA  SYSTEM CONROL N + P (ref. 00678)",
      "product_quantity": "5,437",
      "reason_for_recall": "A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.",
      "recall_initiation_date": "20191118",
      "center_classification_date": "20191223",
      "termination_date": "20230112",
      "report_date": "20200101",
      "code_info": "REF: 00678  Lot Numbers: 253145, 253416, 253655, 254040, 254206, 254401, 254573, 254760    UDIs: (01)03607450006780(11)171130(17)191130(10)253145(241)00678  (01)03607450006780(11)180131(17)200131(10)253416(241)00678  (01)03607450006780(11)180331(17)200331(10)253655(241)00678  (01)03607450006780(11)180630(17)200630(10)254040(241)00678  (01)03607450006780(11)180731(17)200731(10)254206(241)00678  (01)03607450006780(11)180930(17)200930(10)254401(241)00678  (01)03607450006780(11)181031(17)201031(10)254573(241)00678  (01)03607450006780(11)181130(17)201130(10)254760(241)00678",
      "more_code_info": ""
    }
  ]
}