  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89750",
      "recalling_firm": "Philips Healthcare",
      "address_1": "3000 Minuteman Rd",
      "address_2": "",
      "postal_code": "01810-1032",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0737-2022",
      "product_description": "Philips StentBoost Live R2.0 application, when used with the  Philips Azurion with software release R2.x",
      "product_quantity": "42 Systems",
      "reason_for_recall": "Due to an incorrect configuration setting in the EPX database of StentBoost Live, acquisition does not stop automatically as it should. When the user selects the StentBoost Live protocol on the  Philips Azurion system, the following on-screen message is displayed to the user in the Philips StentBoost Live application: \"Press the cine pedal until the acquisition stops\". However, due to the incorrect configuration of the EPX database, the acquisition does not stop after 40 images, but instead it continues as long as the pedal is pressed.",
      "recall_initiation_date": "20220202",
      "center_classification_date": "20220405",
      "termination_date": "20241105",
      "report_date": "20220413",
      "code_info": "Azurion R2.1",
      "more_code_info": ""
    }
  ]
}