  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Covington",
      "address_1": "8195 Industrial Blvd NE",
      "reason_for_recall": "Package Defect: Products may be at risk for having a slit defect on the package and that may affect the product sterility.",
      "address_2": "",
      "product_quantity": "452,728 units",
      "code_info": "Product Code: 0038460, Lot Numbers: NGZC3768, NGZC5029, NGZD0924, NGZD1664, NGZD1983, NGZD3103, NGZD3241, NGZD3989, NGZD4053, GZD4923, NGZE1190, NGZE2134, NGZE3406, NGZF0305, NGZF1546, NGZF2589, NGZF3934, NGZF4488 and NGZF4885.",
      "center_classification_date": "20161212",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Internationally to Australia, New Zealand, Canada, Mexico, France and Italy.",
      "state": "GA",
      "product_description": "BARD¿ Irrigation, Syringe 70cc With Resectoscope Tip, Catheter Tip, Luer Adapter And Cap, Rx Only, Sterile,    The BARD¿ Syringe is available in a single use plastic design which features a 60cc or 70cc capacity graduated barrel. It has a thumb ring plunger for one-handed use. A catheter adapter tip and Luer adapter tip are included. BARD¿ Syringe allows one-hand use with a 60cc or 70cc graduated barrel and thumb-ring plunger and is indicated for catheter irrigation use. BARD¿ Syringe is intended for Cytoscope or Resectoscope irrigation, catheter irrigation and Foley catheter balloon inflation.",
      "report_date": "20161221",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "C.R. Bard, Inc.",
      "recall_number": "Z-0735-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75469",
      "termination_date": "20211005",
      "more_code_info": "",
      "recall_initiation_date": "20160803",
      "postal_code": "30014-1497",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}