{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Oro Valley",
      "state": "AZ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78942",
      "recalling_firm": "Ventana Medical Systems Inc",
      "address_1": "1910 E Innovation Park Dr",
      "address_2": "N/A",
      "postal_code": "85755-1962",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,",
      "recall_number": "Z-0734-2018",
      "product_description": "CINtec PLUS Cytology Kit (CE-IVD), Ventana Part Number 605-100, Roche GMMI 06889565001  in vitro diagnostic",
      "product_quantity": "748 units",
      "reason_for_recall": "Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.",
      "recall_initiation_date": "20171219",
      "center_classification_date": "20180914",
      "termination_date": "20211122",
      "report_date": "20180307",
      "code_info": "UDI 04015630972807, Lot Numbers: Y14122, Y18107"
    }
  ]
}