  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Munich",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89662",
      "recalling_firm": "Brainlab AG",
      "address_1": "Olof-palme-str. 9",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Distribution was made to AZ, AR, CA, CO, FL, GA, IN, KY, MD, MS, MO, NJ, NY, NC, OK, OR, PA, TX, VA, and WA.  There was no government or military distribution.  Foreign distribution was made to Australia, Austria, Bahrain, Belgium, China, Denmark, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, New Zealand, Poland, Singapore, Slovakia, South Korea, Spain, Thailand, Turkey, United Kingdom, and United Arab Emirates.",
      "recall_number": "Z-0733-2022",
      "product_description": "ExacTrac Dynamic software model numbers:    20910-01, ETD Positioning and Moni. 1.0.0 INST;  20910-01A, ETD Positioning and Moni. 1.0.1 INST;  20910-01B, ETD Positioning and Moni. 1.0.2 INST; and   20910-01C ETD Positioning and Moni. 1.0.3 INST.",
      "product_quantity": "205 systems",
      "reason_for_recall": "The yaw angle may be incorrect for CBCT positioning workflows using setup beams with Varian LINACs.",
      "recall_initiation_date": "20220225",
      "center_classification_date": "20220308",
      "report_date": "20220316",
      "code_info": "Software versions 1.0.0, 1.0.1, 1.0.2, 1.0.3, GTIN 04056481142315."
    }
  ]
}