      "country": "United States",
      "city": "Fond Du Lac",
      "address_1": "336 Trowbridge Dr",
      "reason_for_recall": "GF Health Products, Inc., in conjunction with OKIN America, has initiated a voluntary field correction affecting the OKIN America \"DUOMAT 3\" Head/Foot motor used in the manufacture of our Lumex Patriot Full Electric Bed Model US0458/US0468 and Patriot LX Full Electric Bed Model US6000.  The failure mode is directly related to contaminated material in the plastic housing used to produce the \u001cDUOMAT 3\u001d motor. Contaminated motor housings may crack resulting in a motor that will no longer actuate the head section per the intended design.  If the head section were to fall it is unlikely but possible an injury could occur. No injuries have been reported.",
      "address_2": "",
      "product_quantity": "3,547 (3063 beds, 484 motors)",
      "code_info": "BED                                                      SERIAL NUMBER RANGE  US0458 / US0468 Patriot Full Electric 120V  697142145108 through 697142147927  US0458 / US0468 Patriot Full Electric 220V  697144140132 through 697144140172  US6000 Patriot LX Full Electric 120V             69204141262 through 69204141737      SERVICE MOTOR KITS WITH SHIP DATES OF MAY 4, 2015 THROUGH JUNE 30, 2016  Description                                                                           Part Number  Motor and Pendant Patriot Full Electric 120V               690-3001-943 / 690-7001-943  Boxed Full Electric 220V Motor                                     690-3220-943  Motor and Pendant Patriot LX Full Electric 120V          690-8004-912",
      "center_classification_date": "20161212",
      "distribution_pattern": "Worldwide Distribution - USA ( nationwide) Distribution including Puerto Rico  and to the countries of : Bahamas, Canada, Chile, El Salvador and Santo Domingo.",
      "state": "WI",
      "product_description": "Lumex, Patriot Full Electric Homecare Bed, & Patriot LX Full Electric Homecare Bed containing the Okin Dumat 3 Head/Foot Motors.      Description                                                                           Part Number  Motor and Pendant Patriot Full Electric 120V               690-3001-943 / 690-7001-943  Boxed Full Electric 220V Motor                                     690-3220-943  Motor and Pendant Patriot LX Full Electric 120V          690-8004-912",
      "report_date": "20161221",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GF Health Products, Inc",
      "recall_number": "Z-0732-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "75460",
      "termination_date": "20190429",
      "more_code_info": "",
      "recall_initiation_date": "20160930",
      "postal_code": "54937-9103",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}