{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Oro Valley",
      "state": "AZ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78942",
      "recalling_firm": "Ventana Medical Systems Inc",
      "address_1": "1910 E Innovation Park Dr",
      "address_2": "N/A",
      "postal_code": "85755-1962",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,",
      "recall_number": "Z-0731-2018",
      "product_description": "iView DAB IHC Detection Kit, Ventana part Number 760-091, Roche GMMI 05266157001 in vitro diagnostic",
      "product_quantity": "1980 units",
      "reason_for_recall": "Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.",
      "recall_initiation_date": "20171219",
      "center_classification_date": "20180914",
      "termination_date": "20211122",
      "report_date": "20180307",
      "code_info": "UDI 04015630971749, Lot Numbers:  Y11834, Y24245"
    }
  ]
}