      "country": "United States",
      "city": "Hartland",
      "address_1": "900 Walnut Ridge Dr",
      "reason_for_recall": "Reporting feature times out after inactivity for more than an hour sending the user back to the study list, which causes all reporting data being entered to be lost.",
      "address_2": "",
      "product_quantity": "116 sites potentially have the effected versions",
      "code_info": "Versions V10.1 and 10.1.1",
      "center_classification_date": "20161209",
      "distribution_pattern": "Nationwide Distribution",
      "state": "WI",
      "product_description": "Merge Cardio software.     Merge Cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information from Camtronics and other vendors systems including images, hemodynamic studies and reports, measurements (via import from DICOM Structured reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data",
      "report_date": "20161221",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Merge Healthcare, Inc.",
      "recall_number": "Z-0730-2017",
      "initial_firm_notification": "Other",
      "product_type": "Devices",
      "event_id": "75714",
      "termination_date": "20180323",
      "more_code_info": "",
      "recall_initiation_date": "20151001",
      "postal_code": "53029-8347",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}