  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69838",
      "recalling_firm": "Stryker Endoscopy",
      "address_1": "5900 Optical Ct",
      "address_2": "N/A",
      "postal_code": "95138-1400",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.",
      "recall_number": "Z-0727-2015",
      "product_description": "PKG, 5MM INSERT, ENDO GRASPER, 33CM, P/N 0250080699  Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.",
      "product_quantity": "180,573 units total",
      "reason_for_recall": "The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods",
      "recall_initiation_date": "20141117",
      "center_classification_date": "20141217",
      "termination_date": "20150717",
      "report_date": "20141224",
      "code_info": "844411  844628  844811  846760  940456  943679  946279  1042100  1043033  1043101  1045370  1140297  1141155  1142927  1144484  1145067  1145725  1146838  1242927  1244166  1245602  1340116  1341541  1342757  1342758  1344780  1346663  544589"
    }
  ]
}