  "results": [
    {
      "status": "Ongoing",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89375",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "address_2": "N/A",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide  Foreign:  Argentina  Austria  Bangladesh  Brazil  Chile  China  Denmark  Dominican Republic  France  Germany  Greece  India  Indonesia  Iran  Israel  Italy  Japan  Korea, Republic of  Malaysia  Mexico  Netherlands  Philippines  Russian Federation  Saudi Arabia  South Africa  Spain  Sri Lanka  Switzerland  Taiwan  Thailand  Turkey  United Kingdom  Virgin Islands (U.S.)",
      "recall_number": "Z-0725-2022",
      "product_description": "Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chiller without an extended drip tray:  Model Number: 722031",
      "product_quantity": "1 unit",
      "reason_for_recall": "Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.",
      "recall_initiation_date": "20211124",
      "center_classification_date": "20220304",
      "report_date": "20220316",
      "code_info": "S/N:  OUS  109"
    }
  ]
}