{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89375",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "address_2": "N/A",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide  Foreign:  Argentina  Austria  Bangladesh  Brazil  Chile  China  Denmark  Dominican Republic  France  Germany  Greece  India  Indonesia  Iran  Israel  Italy  Japan  Korea, Republic of  Malaysia  Mexico  Netherlands  Philippines  Russian Federation  Saudi Arabia  South Africa  Spain  Sri Lanka  Switzerland  Taiwan  Thailand  Turkey  United Kingdom  Virgin Islands (U.S.)",
      "recall_number": "Z-0724-2022",
      "product_description": "Philips Allura Xper, UNIQ(Allura Xper system-China)  Interventional Fluoroscopic X-ray System products with a Laird chiller WITHOUT an extended drip tray:  Model Number: 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722029, 722033, 722034, 722036, 722038, 722039.    Allura Xper F010 {C+F). 722003  Allura Xper F010/10 722005  Allura Xper F020 722006  Allura Xper F020/10 and F020/20 722008  Allura Xper F010 722010  Altura Xper F010/10 722011  Allura Xper F020 722012  Allura Xper FD20 biplane 722013  Allura Xper FDlO OR Table 722014  Allura Xper FD20 OR Table 722015  Allura Xper FDl0/10 OR Table 722019  Allura Xper FD20 Biplane OR Table 722020  Allura Xper FD10 OR Table 722022  Allura Xper FD20 OR Table 722023  Allura Xper FD10/10 OR Table 722024  Altura Xper FD20 Biplane OR Table 722025  Allura Xper FD10 722026  Allura Xper  FD10/10 722027  Altura Xper FD20/10 biplane 722029  Allura Xper FD10 OR Table 722033  Allura Xper FD10/10 OR Table 722034  Allura Xper FD20/10 OR Table 722036  Allure Xper F020/20 722038  Altura Xper FD20/20 biplane OR Table 722039",
      "product_quantity": "USA: 63 systems\tR.O.W.: 127 systems",
      "reason_for_recall": "Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.",
      "recall_initiation_date": "20211124",
      "center_classification_date": "20220304",
      "report_date": "20220316",
      "code_info": "Allura Xper FD10 722003/722026   (US)  S/N:   24  40  71  162  191  266  283  294  319  347  406  485  825  959  965  967  985  990  1246  1353  1444  1944  1949  UDI:  (01)00884838054189(21)825  (01)00884838054189(21)967  (01)00884838054189(21)1246  (01)00884838054189(21)965  (01)00884838054189(21)959  (01)00884838054189(21)990  (01)00884838054189(21)985      Allura Xper FD10/10 722005  S/N: 144 US  OUS:  37  52  74  79  80  108  279  417  459    Allura Xper FD20   722006  S/N: US  44  72  234  246  263  268  327  333  429  469  471  498  514  535  550  554  588  604  718  754  766  861  1174  1274  1355  1400  1783  1831  1844  1878  1886  1931  2018  UDI: (01)00884838054202(21)893    OUS:  11  40  89  102  112  125  151  244  364  408  431  484  568  580  594  597  602  640  643  704  811  940  971  1262  1312  1402  1405  1424  1465  1481  1658  1691  1794  1796  1815  1852  1859  1879  1891  1948  1959  1974  1985  1988  2141  2311  2521  2554      Allura Xper FD20 Biplane 722008  S/N: 33, 82, 374 (US)  OUS:  39  86  224  249  336  384  388  532      Allura Xper FD10  722010  S/N:OUS  163  340  1422  1429  1434    Allura Xper FD20  722012  S/N: OUS  395  463  927  1013  1518  2652  2854  2918    Allura Xper FD20 OR Table  722015  S/N: 30 (OUS)    Allura Xper FD10 722026  S/N OUS:  315  701  748  884  888  943  948  960  971  975  978  984  989  995    Allura Xper FD10/10  722027  S/N: OUS  268  278    Allura Xper FD20/10 722029  S/N:OUS  9    Allura Xper FD10  722026  S/N : US  825  959  965  967  985  990  1246    OUS:  11  40  89  102  112  125  151  244  364  408  431  484  568  580  594  597  602  640  643  704  811  940  971  1262  1312  1402  1405  1424  1465  1481  1658  1691  1794  1796  1815  1852  1859  1879  1891  1948  1959  1974  1985  1988  2141  2311  2521  2554    Allura Xper FD10/10 266 722027  S/N: 266    Allura Xper FD20 722028  S/N: 893  UDI: (01)00884838054202(21)893    Allura Xper FD20/10 722029  S/N: 91, 94 US  UDI: (01)00884838054219(21)94  OUS:   S/N:6"
    }
  ]
}