{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86900",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "N/A",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the countries of Australia, Austria, Bahamas, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cura¿ao,¿St¿Eus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, India, Iran, Ireland, Israel, Italy, Japan, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, P.R. China, Pakistan, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., United Kingdom, Uruguay, Vatican City, Vietnam.  ¿¿¿¿  ¿¿¿¿¿",
      "recall_number": "Z-0721-2021",
      "product_description": "ADVIA Gamma-Glutamyl Transferase (GGT) Reagent - Product Usage: For in vitro diagnostic use in the quantitative determination of gamma-glutamyl transferase in human serum and plasma. Siemens Material Number (SMN):  ADVIA Chemistry: 10309495 (20 mL) 10316298 (70 mL)",
      "product_quantity": "397 units US; 1713 units OUS",
      "reason_for_recall": "Gamma-Glutamyl Transferase (GGT) Reagent   Low End Repeatability-Imprecision. Reagent may demonstrate a coefficient of variation (%CV) outside of the performance data for repeatability (within-run) listed in the Instructions For Use (IFU) on some analyzers.",
      "recall_initiation_date": "20201123",
      "center_classification_date": "20201223",
      "report_date": "20201230",
      "code_info": "All lot codes.  UDI:  10309495 (20 mL) = 00630414508344  10316298 (70 mL) = 00630414209753"
    }
  ]
}