  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69838",
      "recalling_firm": "Stryker Endoscopy",
      "address_1": "5900 Optical Ct",
      "address_2": "N/A",
      "postal_code": "95138-1400",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.",
      "recall_number": "Z-0721-2015",
      "product_description": "PKG, 10MM PEEK MONOPOLAR HANDLE, P/N 0250080618  Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.",
      "product_quantity": "180,573 units total",
      "reason_for_recall": "The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods",
      "recall_initiation_date": "20141117",
      "center_classification_date": "20141217",
      "termination_date": "20150717",
      "report_date": "20141224",
      "code_info": "42167  42169  42172  43238  43239  81309  142647  143476  143477  245841  342652  344115  345130  0442468C  0445112D  0541360D  0541361D  0641114D  0641115D  0644773D  0645377D  0647994D  06CON06071  06CON06126  07022013E  0740507D  0740834D  0741766D  0742442D  0742740D  0744063D  0744684D  0745499D  0746059D  0746497D  0747592D  0840357D  0840678D  0841358D  0842085D  0842130D  0845885D  0846565D  0847512D  9292011  0940124D  0940126D  0940127D  0941052D  0941067D  0942146D  0942277D  0942850D  0943496D  0944142D  0945209D  0945467D  0945715D  1040120D  1040385D  1040583D  1040805D  1041471D  1041627D  1041982D  1042493D  1043025D  1043102D  1043511D  1043778D  1044521D  1044523D  1046109D  1046189D  1046193D  1141605D  1141791D  1142149D  1142927D  1143476D  1143648D  1143972D  1144238D  1144345D  1144346D  1144395D  1144642D  1144656D  1144961D  1145073D  1145344D  1145810D  1145941D  1146537D  1146660D  1146661D  1240125D  1240950D  1241422D  1241512D  1242235D  1242642D  1242713D  1242929  1242929D  1243097  1243097D  1243552  1244644  1340202  1340554  1340554H  1341605  1341605H  1341606  1341606H  1341608  1341608H  1343303  1343303H  1343724  1343724H  1345516  1345516H  1346542  1346542H  1346543  1440947  1440947H  1440955  1440955H"
    }
  ]
}