  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Stockholm",
      "state": "N/A",
      "country": "Sweden",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75633",
      "recalling_firm": "RAYSEARCH LABORATORIES AB",
      "address_1": "Sveavaegen 9",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution in the states of AZ, AR, CA, CT, DE, FL, GA, HI, IL, LA, MA, MD, MN, MI, MS, MO, NJ, NY, NC, OH, PA, RI, TX & WA",
      "recall_number": "Z-0720-2017",
      "product_description": "Radiation Therapy Treatment Planning System     Product Usage:  RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.",
      "product_quantity": "540 units",
      "reason_for_recall": "An error may occur with the display of dose computed on images other than the planning CT (auxiliary CT) when using multiple patient cases in RayStation 5. If a CT image set with the same Frame of Reference as the displayed auxiliary CT set exists in another case, the dose display may be incorrect.  When the error occurs, the display of dose in patient views, including the maximum dose position, the dynamic isodose lines and the dose grid may be incorrect.  The dose value normally displayed in the upper left corner when pointing in the 2D view may be incorrect or missing.",
      "recall_initiation_date": "20161011",
      "center_classification_date": "20161207",
      "termination_date": "20190611",
      "report_date": "20161214",
      "code_info": "Build No: \"5.0.0.37\u001d, \u001c5.0.1.11\u001d or \u001c5.0.2.35\u001d.  Only 5.0.2.35 was distributed in the U.S."
    }
  ]
}