      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72599",
      "recalling_firm": "Philips Electronics North  America Corporation",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) including Puerto Rico and countries of: Albania, Algeria, US Virgin Islands, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Democratic Congo, Costa Rica, Croatia, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.",
      "recall_number": "Z-0716-2016",
      "product_description": "Cardiovascular Allura Centron; Model Number: 722400\t    The Allura CV2O is intended for physicians (e.g. cardiologists and radiologists), assisted by trained hospital staff (e.g. nurses and lab technicians), who are qualified to perform medical procedures on humans (having a maximum weight of 250 kg.) with probable internal diseases or injuries for: \" Dedicated vascular and carotid imaging applications, including diagnostic and interventional procedures. \" Cardiac imaging applications including diagnostics, interventional rocedures, pacemaker implantations and electrophysiology (EP). \" Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.",
      "product_quantity": "13297 in total",
      "reason_for_recall": "The use of Allura Xper, Allura Clarity, Allura CV, Allura Centron and UNIQ systems regarding live images and still images can lead to still images being interpreted as live Images.",
      "recall_initiation_date": "20151103",
      "center_classification_date": "20160130",
      "termination_date": "20200915",
      "report_date": "20160210",
      "code_info": "All serial numbers",
      "more_code_info": ""
    }
  ]
}