  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-05-06",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Nepean",
      "state": "N/A",
      "country": "Canada",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64043",
      "recalling_firm": "Abbott Point of Care Canada Limited",
      "address_1": "185 Corkstown Road",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Worldwide Distribution.",
      "recall_number": "Z-0716-2013",
      "product_description": "Abbott Point of Care i-Stat cTnI cartridge.    The i-Stat cardiac troponin I (cTnI) test is an in vitro diagnostic test for the quantitive measurement of cardiac troponin I (cTnI) in whole blood or plasma. Measurements of cardiac troponin I are used in the diagnosis and treatment of myocardial infarction and as an aid in the risk stratification of patients with acute coronary syndromes with respect to the relative risk of mortality.",
      "product_quantity": "560 boxes (25 cartridges in each box)",
      "reason_for_recall": "APOC has determined that 10 boxes of i-Stat cTnI cartridges from lot P12179 have an incorrect barcode applied to the cartridge portion pack.",
      "recall_initiation_date": "20121101",
      "center_classification_date": "20130124",
      "termination_date": "20130614",
      "report_date": "20130130",
      "code_info": "List number 06F15-03, O6F15-04, 03P90-25, 600-9009-25  Lot number P12179"
    }
  ]
}