  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Glens Falls",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74948",
      "recalling_firm": "Angiodynamics Inc. (Navilyst Medical Inc.)",
      "address_1": "10 Glens Falls Tech Park",
      "address_2": "N/A",
      "postal_code": "12801-3864",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution to CA, FL, GA, HI, LA, MI, MN, MO, NV, OH & TX",
      "recall_number": "Z-0715-2017",
      "product_description": "Peripherally Inserted Central Catheters (PICC), UPN H965750191, Catalog # 75-019     The Maximal Barrier Nursing Kits are Tandem Pack Convenience Kits packaged with a NMI PICC and accessories. The BioFlo PICC with ENDEXO Technology and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring.",
      "product_quantity": "50 cartons (boxes of 5)",
      "reason_for_recall": "There is a potential for stylet stiffening that is not accurately represented on the labeling. The label states that product includes a stylet with an outside diameter (OD) measuring 0.014, however, the product may contain a stylet with an outside diameter (OD) measuring 0.016.",
      "recall_initiation_date": "20160608",
      "center_classification_date": "20161207",
      "termination_date": "20180615",
      "report_date": "20161214",
      "code_info": "Lot #: 4945781"
    }
  ]
}