  "results": [
    {
      "country": "United States",
      "city": "Telford",
      "address_1": "3135 Quarry Rd",
      "reason_for_recall": "After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.",
      "address_2": "",
      "product_quantity": "439",
      "code_info": "with software versions VG2.0.3 and higher.",
      "center_classification_date": "20140110",
      "distribution_pattern": "US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.",
      "state": "PA",
      "product_description": "Draeger Infinity Acute Care System Monitoring Solution",
      "report_date": "20140122",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Draeger Medical Systems, Inc.",
      "recall_number": "Z-0714-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67056",
      "termination_date": "20150520",
      "more_code_info": "",
      "recall_initiation_date": "20131210",
      "postal_code": "18969-1042",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}