  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Westlake",
      "address_1": "810 Sharon Dr",
      "reason_for_recall": "RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains glycerol which can be reduced at the cathode causing a suddenly rising zero current, causing the membrane unit to fail the following Cal2 calibration.",
      "address_2": "",
      "product_quantity": "209 boxes/6 membranes",
      "code_info": "Part #: 942-042, Lot #: R0971",
      "center_classification_date": "20140110",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) including states of: CO, CT, GA, MA, MO & MS and country of: Canada.",
      "state": "OH",
      "product_description": "pO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145  pO2 membrane units, ordering number 942-042 Intended use is in vitro testing of samples of whole blood for pO2 (partial pressure of oxygen).",
      "report_date": "20140122",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Radiometer America Inc",
      "recall_number": "Z-0713-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67044",
      "termination_date": "20140619",
      "more_code_info": "",
      "recall_initiation_date": "20130610",
      "postal_code": "44145-1521",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}