{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "San Clemente",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86426",
      "recalling_firm": "ICU Medical, Inc.",
      "address_1": "951 Calle Amanecer",
      "address_2": "N/A",
      "postal_code": "92673-6212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of  Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.",
      "recall_number": "Z-0711-2021",
      "product_description": "26 IN(66 cm) APPX 3.4ml, 10 DROP BLOOD SET, 170 MICRON FILTER, Spiros, REF: Z2795",
      "product_quantity": "2900",
      "reason_for_recall": "Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.",
      "recall_initiation_date": "20200831",
      "center_classification_date": "20201223",
      "report_date": "20201230",
      "code_info": "Lots: 4758192, 4763891, 4794293, 4807842, 4849283, 4866670, 4877784, 4902534, 4909635                                                                                                          ,"
    }
  ]
}