  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79214",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Distributed in XX states: Ak, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, the District of Columbia, Guam, and Puerto Rico.",
      "recall_number": "Z-0710-2018",
      "product_description": "Artis zee and Artis zeego model Interventional fluoroscopic x-ray   foot switch accessory; Model Numbers: 4787797, 4787805, and 4787813.",
      "product_quantity": "510",
      "reason_for_recall": "The possibility exists for the wireless foot switch to fail due to impacts by external factors, such as electrostatic discharge that exceed a certain intensity. If the foot switch fails, it will no longer be able to be used to release radiation. The hand switch provided can be used to release exposures; however, fluoroscopic examinations will not be possible.",
      "recall_initiation_date": "20170928",
      "center_classification_date": "20180223",
      "termination_date": "20200309",
      "report_date": "20180307",
      "code_info": "Serial Numbers: 10000  10006  10009  10010  10016  10018  10019  10027  10029  10030  10040  10042  10044  10046  10047  10048  10055  10056  10059  10062  10068  10069  10073  10075  10080  10084  10089  10094  10095  10096  10098  10099  10100  10102  10103  10104  10111  10112  10115  10116  10121  10122  10125  10126  10128  10133  1027  1049  1059  1072  1084  1085  1096  1101  1102  1104  1105  1111  1112  1115  1117  1122  1129  1131  1139  1144  1147  1149  1156  1157  1160  1165  1166  1182  1188  1189  1196  1199  1204  1205  1206  1208  1222  1227  1228  1229  1232  1233  1240  1242  1243  1245  1246  1248  1252  1255  1256  1258  1260  1262  1266  1267  1269  1270  1274  1278  1279  1282  1283  1286  1292  1293  1294  1295  1299  1300  13173  15000  15003  15004  15015  15016  15021  15024  15026  15028  15033  15034  15042  15045  15049  15053  15055  15056  15066  15069  15070  15074  15076  15079  15080  15087  15100  15103  15104  15105  15106  15108  15109  15111  15112  15114  15116  15118  15119  15121  15122  15124  15126  15127  15128  15134  15135  15138  15139  15143  15144  15153  15154  15158  15159  15160  15164  15167  15180  15190  15197  15200  15207  15224  17002  17003  17012  17017  17020  17023  17032  17035  17036  17037  17041  17056  17057  17058  17059  17072  17073  17074  17077  17078  17081  17094  17096  17099  17101  17102  17104  17105  17107  17109  17113  17130  17143  17146  17151  17155  17159  17160  17167  17174  17177  20000  20008  20009  20017  20020  20022  20029  20037  20042  20046  20051  20057  20059  20060  20062  20064  20066  20069  20070  20071  20075  20076  20077  20078  20081  20083  20087  20088  20090  20097  20098  20101  20102  20104  20106  20108  20110  20113  20114  20116  20117  20119  20121  20122  20125  20126  20128  20129  20130  20131  20132  20133  20138  20140  20146  20148  20159  20160  20161  20173  20183  20193  20195  20197  20199  20200  20209  20216  20219  20220  20226  20227  20228  20232  20236  20237  20239  20241  20243  20246  20248  20249  20253  20254  20261  20263  20264  20265  20268  20270  20274  20277  20279  20281  20282  20287  20290  20293  20294  20296  20298  20299  20301  20314  20315  20320  20321  20323  20325  20328  20329  20333  20334  20337  20339  20340  20348  20349  20358  20362  20369  20371  20373  20374  20381  20383  20388  20393  20394  20395  20402  20403  20404  20405  20410  20412  20421  20425  20429  20430  20432  20433  20437  20439  20441  20443  20445  20449  20455  20456  20460  20462  20463  20467  20473  20475  20476  20478  20480  20484  20485  20486  20488  20493  20494  20496  20506  20509  20511  20512  20519  20524  20531  20535  20536  20537  20539  20542  20544  20548  20550  20551  20552  20557  20561  20563  20565  20566  20567  20568  20576  20585  20595  20600  20603  20606  20607  20611  20614  20622  20623  20625  20627  20629  20630  20631  20633  20634  20635  20637  20640  20642  20643  20644  20646  20647  20649  20651  20652  20658  20661  20664  20667  20668  20673  20681  20682  20685  20687  20689  20690  20691  20692  20694  20700  20703  20704  20705  20706  20707  20710  20713  20715  20716  20718  20719  20720  20721  20722  20723  20725  20726  20731  20733  20734  20736  20737  20738  20742  20744  20745  20746  20748  20749  20755  20761  20762  20764  20765  20769  20770  20771  20774  20777  20780  20781  20782  20784  20785  20786  20787  20788  20789  20794  20795  20798  2866  2Z003  S25006"
    }
  ]
}