  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Palm Beach Gardens",
      "address_1": "4500 Riverside Drive",
      "reason_for_recall": "During a recent review of documentation Anspach determined that the current directions for use for the MCA cutting burrs and MCA Micro Curved Attachment do not provide clear details regarding the intended use specifically for the MCA burrs.  The MCA burrs are intended for use in delicate bone in otologic procedures such as cochleostomy.  The use of the device in dense bone removal may result in cu",
      "address_2": "",
      "product_quantity": "6,824 imits (4,7324 in USA / 2,100 Foreign)",
      "code_info": "All Lots for   MCA - MICRO CURVED ATTACHMENT,   MCA5-05SD - 0.5MM DIAMOND BALL, 5.6CM,   MCA5-06SD - 0.6MM DIAMOND BALL, 5.6CM,   MCA5-07SD - 0.7MM DIAMOND BALL, 5.6CM,   MCA5-08SD - 0.8MM DIAMOND BALL, 5.6CM,   MCA5-15SB - 1.5MM FLUTED BALL, 5.6CM,   MCA5-15SD - 1.5MM DIAMOND BALL, 5.6CM,   MCA5-1SB - lMM FLUTED BALL, 5.6CM,   MCA5-1SD - lMM DIAMOND BALL, 5.6CM,   MCA5-2SB - 2MM FLUTED BALL, 5.6CM,   MCA5-2SD - 2MM DIAMOND BALL, 5.6CM,   MCA7-05SD - 0.5MM DIAMOND BALL, 7.3CM,   MCA7-06SD - 0.6MM DIAMOND BALL, 7 .3CM,   MCA7-07SD - 0.7MM DIAMOND BALL, 7.3CM,   MCA7-08SD - 0.8MM DIAMOND BALL, 7.3CM,   MCA7-15SB - 1.5MM FLUTED BALL, 7.3CM,   MCA7-15SD \u0013 1.5MM DIAMOND BALL, 7.3CM,   MCA7-1SB - lMM FLUTED BALL, 7.3CM,    MCA7-1SD -  1 MM DIAMOND BALL, 7 .3CM,    MCA7-2SB - 2MM FLUTED BALL, 7.3CM and   MCA7-2SD - 2MM DIAMOND BALL, 7.3CM.",
      "center_classification_date": "20140109",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Canada, Netherlands, Great Britain, France, Sweeden, Italy, Spain, Japan, South Africa, Australia, Switzerland,",
      "state": "FL",
      "product_description": "Multiples Anspach Micro Curved Attachments (MCA) and Curved Burr Support S leves (Cutters and Attachments)  MCA - MICRO CURVED ATTACHMENT,   MCA5-05SD - 0.5MM DIAMOND BALL, 5.6CM,   MCA5-06SD - 0.6MM DIAMOND BALL, 5.6CM,   MCA5-07SD - 0.7MM DIAMOND BALL, 5.6CM,   MCA5-08SD - 0.8MM DIAMOND BALL, 5.6CM,   MCA5-15SB - 1.5MM FLUTED BALL, 5.6CM,   MCA5-15SD - 1.5MM DIAMOND BALL, 5.6CM,   MCA5-1SB - lMM FLUTED BALL, 5.6CM,   MCA5-1SD - lMM DIAMOND BALL, 5.6CM,   MCA5-2SB - 2MM FLUTED BALL, 5.6CM,   MCA5-2SD - 2MM DIAMOND BALL, 5.6CM,   MCA7-05SD - 0.5MM DIAMOND BALL, 7.3CM,   MCA7-06SD - 0.6MM DIAMOND BALL, 7 .3CM,   MCA7-07SD - 0.7MM DIAMOND BALL, 7.3CM,   MCA7-08SD - 0.8MM DIAMOND BALL, 7.3CM,   MCA7-15SB - 1.5MM FLUTED BALL, 7.3CM,   MCA7-15SD \u0013 1.5MM DIAMOND BALL, 7.3CM,   MCA7-1SB - lMM FLUTED BALL, 7.3CM,    MCA7-1SD -  1 MM DIAMOND BALL, 7 .3CM,    MCA7-2SB - 2MM FLUTED BALL, 7.3CM and   MCA7-2SD - 2MM DIAMOND BALL, 7.3CM.    The MCA attachment is a reusable device used with dissection tools. Dissection tools are single use, disposable cutter designed for cutting and shaping delicate bone, primary in otology procedures for delicate bone cutting.",
      "report_date": "20140115",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "The Anspach Effort, Inc.",
      "recall_number": "Z-0708-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67124",
      "termination_date": "20140623",
      "more_code_info": "",
      "recall_initiation_date": "20121026",
      "postal_code": "33410-4235",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}