{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Raynham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63665",
      "recalling_firm": "DePuy Spine, Inc.",
      "address_1": "325 Paramount Dr",
      "address_2": "N/A",
      "postal_code": "02767-5199",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) including the states of: AZ, CA, FL, HI,  IL, MD,MI,  MO,NC, TN, TX and the country of Australia.",
      "recall_number": "Z-0708-2013",
      "product_description": "DePuy Spine  VIPER POLY SCREW 5.5  9 X70MM Tl  Product Code: 1867-15-970          Spinal Implant Component.",
      "product_quantity": "24 units",
      "reason_for_recall": "Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head",
      "recall_initiation_date": "20121107",
      "center_classification_date": "20130123",
      "termination_date": "20131224",
      "report_date": "20130130",
      "code_info": "All lots"
    }
  ]
}