{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Louisville",
      "address_1": "826 Coal Creek Cir",
      "reason_for_recall": "Under certain circumstances, the percutaneous pin adapter used with the StealthAir Frame Assembly may slightly rotate after surgical placement and not return to its original position, even when connections are tight and secure.",
      "address_2": "",
      "product_quantity": "1,297 percutaneous pin adapters",
      "code_info": "Percutaneous pin adapter lot numbers:   150114, 150505, 150506, 150904, 151005, 151007 160201, 160224, 160419, 160617, 160707, 170216, 180110, and 180409.",
      "center_classification_date": "20191219",
      "distribution_pattern": "Distribution was nationwide.  There was government/military/foreign distribution.",
      "state": "CO",
      "product_description": "Medtronic Adapter Percutaneous Pin, REF 9734752 (GTIN 00643169529489)  and REF 9734752-G02 (GTIN 00763000054915), contained in the following kits:  Upgrade Set Percutaneous Pin, REF 9735502 (GTIN 00643169529533) and REF 9735502-G02 (GTIN 00763000208462); Medtronic Adapter Percutaneous Pin Kit, REF 9734752K (GTIN 00643169540521); and Medtronic Adapter Percutaneous Pin Kit, REF 9734752K (GTIN 00643169641006) (Japan).  The percutaneous pin adapter is a component of the StealthAir Frame Assembly.",
      "report_date": "20191225",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medtronic Navigation, Inc.",
      "recall_number": "Z-0707-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "84045",
      "termination_date": "20201217",
      "more_code_info": "",
      "recall_initiation_date": "20181206",
      "postal_code": "80027-9710",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}