      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79139",
      "recalling_firm": "Philips Electronics North  America Corporation",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) Distribution,  Internationally to Canada.",
      "recall_number": "Z-0704-2018",
      "product_description": "digital x-ray detector    ProGrade R1 - solid state X ray imager (flat panel/digital imager)    As a part of a radiographic system, the Philips ProGrade is intended to acquire, process, store, display, and export digital radiographic images. The Philips ProGrade is suitable for all routine radiographic examinations, including specialist area like intensive care, trauma, or pediatric work, excluding mammography. United States only: The Eleva Workspot is not intended for fluoroscopy and angiography.",
      "product_quantity": "157",
      "reason_for_recall": "It the WiFi connection between the SkyPlate detector and HP transfer point is weak, an image may fail to transfer from the SkyPlate detector to the system.  The image remains in the memory of the detector, but cannot be transferred wirelessly or by use of the backup cable. To continue, the operator can reset the SkyPlate detector by removing its batteries, but the acquired image is lost and a  re-take is necessary.",
      "recall_initiation_date": "20171211",
      "center_classification_date": "20180223",
      "termination_date": "20200529",
      "report_date": "20180307",
      "code_info": "ProGrade R1,  solid state X ray imager, 712090  Serial #s:  SN17000011  SN15000026  SN17000022  SN15000038  SN16000001  SN14000035  SN17000020  SN17000003  SN17000008  SN17000013  SN14000024  SN14000034  SN15000012  SN15000024  SN15000022  SN16000035  SN17000018  SN17000024  SN15000049  SN16000006  SN16000010  SN17000001  SN17000023  SN15000042  SN16000011  SN16000020  SN14000007  SN17000025  SN15000051  SN15000028  SN17000014  SN15000016  SN15000035  SN17000015  SN16000041  SN16000022  SN16000007  SN15000005  SN16000009  SN16000040  SN15000046  SN16000016  SN16000015  SN14000013  SN15000006  SN17000030  SN16000012  SN16000014  SN15000039  SN16000039  SN14000028  SN15000032  SN16000042  SN17000021  SN15000013  SN17000007  SN15000047  SN16000024  SN17000002  SN17000005  SN14000022  SN17000004  SN16000030  SN16000031  SN16000032  SN14000006  SN15000033  SN16000021  SN16000029  SN14000016  SN15000031  SN14000003  SN14000014  SN15000019  SN14000023  SN16000019  SN16000038  SN16000017  SN16000004  SN14000017  SN15000007  SN16000036  SN16000037  SN15000029  SN15000050  SN15000009  SN17000010  SN14000029  SN14000032  SN15000053  SN14000015  SN16000034  SN14000001  SN15000004  SN15000014  SN16000002  SN14000026  SN14000005  SN14000021  SN14000019  SN15000045  SN15000030  SN15000036  SN15000025  SN14000033  SN15000001  SN15000002  SN16000018  SN17000012  SN17000016  SN14000020  SN15000043  SN16000033  SN14000008  SN16000027  SN14000031  SN15000003  SN15000008  SN15000017  SN15000027  SN14000025  SN15000044  SN16000023  SN14000004  SN15000041  SN16000028  SN14000010  SN14000018  SN16000013  SN14000027  SN16000043  SN15000015  SN15000011  SN15000040  SN16000003  SN14000011  SN14000012  SN14000009  SN15000037  SN16000005  SN15000020  SN15000023  SN14000030  SN15000018  SN17000006  SN16000026  SN15000034  SN17000026  SN15000052  SN15000010  SN15000048  SN15000054  SN15000055  SN16000008  SN17000009  SN16000025  SN15000021"
    }
  ]
}