  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Saint Paul",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75644",
      "recalling_firm": "3M Company - Health Care Business",
      "address_1": "3m Center 2510 Conway Ave , B# 275-5-W-6",
      "address_2": "N/A",
      "postal_code": "55144-0001",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of AR, AZ, CA, FL, GA, ID, IL, KY, NC, NY, TN, TX, VA and the countries of Canada, Chile, Costa Rica, Germany, Japan, Mexico and South Africa.",
      "recall_number": "Z-0701-2017",
      "product_description": "3M Universal Electrosurgical Pad, REF 9130 (100 pads / case) and    3M Universal Electrosurgical Pad with Cord, REF 9135-LP (40 pads / case)    Product Usage:  3M Universal Electrosurgical Pads are designed to work with most electrosurgical unite (ESU) for virtually every surgical application where electrosurgery is utilized to provide a safe return path for electrosurgical current. Solid Universal Electrosurgical Pads are for use with generators that do not have a Contact Quality Monitoring System (CQMS). Split Style Universal Electrosurgical Pad is for use with ESUs that have a CQMS (i.e. REM, ARM, NESSY, etc.). 3M Universal Electrosurgical Pads are designed to be used on any patient where full skin contact and a suitable placement site can be obtained. There are no patient weight restrictions for use of this product. Use of this product for unintended application could lead to an unsafe condition.",
      "product_quantity": "32,000 pads",
      "reason_for_recall": "3M is recalling Universal Electrosurgical Pads because a report of an unintended material (process liner) in the product.  This may prevent or impede the safe return of electrosurgical current following Electrosurgical Unit (ESU) activation.  This defect has the potential to increase the risk of a patient burn directly under the pad or at an alternate site of the body.",
      "recall_initiation_date": "20161026",
      "center_classification_date": "20161203",
      "termination_date": "20170801",
      "report_date": "20161214",
      "code_info": "Product REF 9130 Lot Numbers:  2019-02DT (Expiration 02-2019) and 2019-03DU (Expiration   03-2019)  Product REF 9735-LP Lot Numbers:  2019-02DB and 2019-02CY.  (Both with  Expiration 02-2019)"
    }
  ]
}