  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72864",
      "recalling_firm": "Mindray DS USA, Inc. dba Mindray North America",
      "address_1": "800 Macarthur Blvd",
      "address_2": "N/A",
      "postal_code": "07430-2001",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution.",
      "recall_number": "Z-0701-2016",
      "product_description": "Panorama Patient Monitoring Network.    Software Kit 0020-00-0205-19A, View Station Software Kit 0020-00-0206-16A, Work Station Software Kit 0020-00-0207-16A, Software Kit Upgrade TECH-00-910.    The Panaroma Patient Monitoring Network is intended for use in a fixed location in the healthcare facility setting as a central viewing station. It is not intended to be directly connected to patient at any time or installed in a patient's vicinity.",
      "product_quantity": "63 products",
      "reason_for_recall": "An issue with the Panorama Central Station may cause the system to spontaneously restart. This may occur after about 49 days of continuous operation.  On restart, monitoring will be lost for about 2.5 minutes. Settings and configurations are maintained.",
      "recall_initiation_date": "20151215",
      "center_classification_date": "20160129",
      "termination_date": "20170329",
      "report_date": "20160210",
      "code_info": "version 11.5.1",
      "more_code_info": ""
    }
  ]
}