  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75458",
      "recalling_firm": "Medtronic Neuromodulation",
      "address_1": "7000 Central Ave",
      "address_2": "N/A",
      "postal_code": "55421-1241",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide including DC    OUS: Canada and others.",
      "recall_number": "Z-0700-2017",
      "product_description": "Model 37751 Recharger    Product Usage:  The Medtronic 37751 Recharger is a hand-held device used to charge the battery in a patient s neurostimulator. It includes a display to provide information on the charging system. The charging system consists of the Model 37751 Recharger, an AC power supply and power cord (Model 37761) and an antenna.",
      "product_quantity": "42,887",
      "reason_for_recall": "Medtronic has identified an increased number of complaints from customers involving reports of Rechargers  (Medtronic Model 37751 Recharger, which is included in Models 37754 and 97754 Charging Systems,Spinal Cord Stimulation)that are in an unresponsive error state, where the Recharger is non-functional with a blank display screen and is beeping every 5 seconds. Medtronic has determined all Rechargers manufactured starting in November 2014 (indicated by serial numbers beginning with \u001cNKA4\u001d or \u001cNKU4\u001d) are more susceptible to this error state.",
      "recall_initiation_date": "20161003",
      "center_classification_date": "20161203",
      "termination_date": "20220504",
      "report_date": "20161214",
      "code_info": "serial numbers beginning with \u001cNKA4\u001d or \u001cNKU4"
    }
  ]
}