      "country": "United States",
      "city": "Lawrence",
      "address_1": "350 Merrimack St",
      "reason_for_recall": "There is a potential patient health risk while performing chronic hemodialysis treatments in the home setting with  for patients with low blood volume (2.4 Liters or less in a patient with body weight of 30kg or less) when prescribing larger dialysate volumes.",
      "address_2": "",
      "product_quantity": "1803 units",
      "code_info": "Product 1) UDI # - M535NX100010  Product 2) UDI # - M535NX1000160 and M535NX100016A0  Product 3) UDI # - M535NX100030 and M535NX10003A0  Product 4) UDI # - M535NX1000100 and M535NX100010A0     Lot Numbers:   60677021, 60877053, 70477033, 71177020, 80377002, 80477018, 80877011, 81077018, 81177047, 90277048 and 90677002",
      "center_classification_date": "20191218",
      "distribution_pattern": "AL, AZ, CA, NC & VA",
      "state": "MA",
      "product_description": "Low Volume Cartridge IFU (SKU CAR-125-B) used with the following devices:    1) NxStage One, Model NX1000-1;  2) NxStage VersiHD Cycler, Model Numbers: NX1000-16, and NX1000-16-A;  3) NxStage One, Model Numbers: NX1000-3 and NX1000-3-A;  4) NxStage One, Model Numbers: NX1000-10 and NX1000-10-A;",
      "report_date": "20191225",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "NxStage Medical, Inc.",
      "recall_number": "Z-0697-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "83706",
      "termination_date": "20210201",
      "more_code_info": "",
      "recall_initiation_date": "20190827",
      "postal_code": "01843-1748",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}